Clamp-On Ultrasonic Flowmeters Ensure Reliable Measurement of Hot Liquids with Solids

Summary

Zhejiang Xinhecheng Co., Ltd., founded in 1999, manufactures functional chemicals including nutritional ingredients, fragrances and polymer materials. For a trial project, the customer needed to measure a liquid containing 10% metallic solid particles at an operating temperature of 150°C. Traditional electromagnetic and differential-pressure flowmeters proved unsuitable because electrodes and primary elements were quickly worn down by the solid particles, requiring frequent maintenance, causing unplanned shutdowns and reduced yield. Clamp-on ultrasonic technology stood out because it can be installed externally without modifying the pipe or interrupting the process.

Application

Fig. 1: C-PT 1 MHz clamp-on transducer installed on a 2″ stainless steel line carrying the hot liquid catalyst.

MediumLiquid catalyst
Pipe size and thicknessDN50 (2″)
Pipe materialSS 304
Pressure0.2 MPa (29 psig)
Flow rate5 – 30 m³/h (22 – 132 GPM)
TemperatureUp to 175°C (347°F)
Requested accuracy±1-2% of reading
Panametrics

Benefits:

  • Low to zero maintenance
  • No pressure drop
  • No process interruption
  • High reliability and long term stability

Challenge

Interrupting production was not feasible, so the operator required a solution that could be implemented without disrupting ongoing operations. As there was no prior experience with this specific application, a field test was conducted to confirm that the flowmeter could perform under the process conditions. Furthermore, the customer expected the clamp-on ultrasonic liquid flowmeter to require minimal maintenance by design.

Solution

Panametrics conducted the field test using a PT900 portable flowmeter with 1 MHz clamp-on transducers. The transducers were mounted on a vertical pipe (Fig. 1) with flow from bottom to top. As the pipe was insulated, the insulation was removed for the test, and installation took approximately 15 minutes. The customer assessed the meter accuracy by triggering the control valve and comparing the resulting changes in flow – the flowmeter responded as expected, with readings tracking the valve’s percentage opening. Satisfied with the results, the customer decided to purchase permanent clamp-on instrumentation: the XMT1000LC with C-PT 1 MHz transducers, with an initial plan for two units. The meters have now been in service for several years, continuing to meet expectations while reducing maintenance costs and improving production yield. For future installations, the customer plans to use C RS HT transducers rated up to 230°C.

C-PT 1 MHz clamp-on transducer installed on insulated 2-inch pipe

Panametrics PanaFlow LC clamp-on ultrasonic liquid flowmeter

Datasheet (PDF) Panametrics PanaFlow LC

Source: Panametrics Application Note PANA148AN (panametrics.com) — © Panametrics LLC. Localized and prepared by INTECH CONTROL, official Panametrics representative for Slovakia and Hungary.