by Bohuslav Bobrik Bohuslav Bobrik
APPLICATION NOTE
Chemical plant at dusk – oxygen monitoring in PVC production

Oxygen Monitoring of VCM in PVC Production

PVC (polyvinyl chloride) ranks as the third most widely produced synthetic plastic polymer, after polyethylene and polypropylene, with roughly 40 million tons manufactured worldwide every year. It is used for pipes, vinyl flooring and siding, hoses, cable coatings, medical devices, plumbing and automotive parts. In PVC production, oxygen monitoring is essential for product quality, yield and emissions.

Application

Vinyl chloride monomer (VCM) is an intermediate in PVC production. Because it is harmful to the environment, VCM emissions should be kept as low as possible. Residual VCM is stripped from the PVC product with steam, and this stage requires careful oxygen measurement: air entering the process leads to excess VCM and lower yield. By monitoring the oxygen concentration in the VCM gas, operators can adjust the oxygen level on demand and keep it to a minimum while using the least possible amount of nitrogen.

Electrolysis of salt water produces chlorine, caustic soda and hydrogen. Ethylene and chlorine are combined into ethylene dichloride (EDC), which is cracked to vinyl chloride – the basic building block of PVC. Polymerization then links the VCM molecules into long PVC chains, producing a white thermoplastic powder.

ApplicationOxygen monitoring during stripping of residual VCM in PVC production
Measuring range0–21 % O2
Typical concentration2–4 % O2 in N2/VCM
Sample temperature40 °C (heated sample system)
PressureAmbient
Outputs4–20 mA analog, RS232 digital interface
Display / controllerProgrammable relays for automatic calibration
Sample systemVacuum pump or eductor, liquid trap, hazardous area certification
Panametrics

Benefits:

  • Rugged and reliable, with no moving parts
  • Sensor is not consumed, unlike electrochemical cells
  • Simple field calibration
  • Turnkey sample conditioning system
Panametrics XMO2pro oxygen transmitter

Challenge

Attempts to measure oxygen in this process with fuel cells have proved problematic and inconsistent because of the high acid content and condensate. The cells degrade within a period that makes them uneconomical and maintenance-intensive.

Solution

The Panametrics XMO2pro thermoparamagnetic oxygen transmitter is well suited to this application. It has no moving parts and is stable and rugged enough to operate in the corrosive VCM gas.

The XMO2pro is delivered with a sample system equipped with a liquid trap and either a vacuum pump or an eductor. The turnkey system also calibrates the transmitter automatically, using nitrogen as the zero gas and instrument air (21 % oxygen) as the span gas.

Panametrics XMO2pro Oxygen Analyzer

Datasheet (PDF) Panametrics XMO2pro